- CBC: WBC, RBC, HGB, HCT, MCV, MCH, MCHC, RDW, RDW-SD, PLT, MPV
- Differential: NE, LY, MO, EO, BA, NRBC, NE#, LY#, MO#, EO#, BA#, NRBC#
- Reticulocytes: RET, RET#, MRV, IRF
- Monocyte Distribution Width (MDW) (when enabled)
- Body fluids (spinal, serous and synovial): TNC, RBC, Pleural, Pericardial, Peritoneal
|
|
|
Up to 100 samples/hour |
|
|
4 required reagents; environmentally friendly methanol- and ethylene glycol-free |
|
|
165 μL aspiration volume; 0.5 ml dead volume |
|
|
- Redundant power computers in a workcell
- 90–264 VAC and 48–62 Hz
- DxH 900 SPM: 520 W
- DxH standard computer: 160 W
- Monitor: 35 W
|
|
|
| Parameter |
Units |
Analytical Measuring Range |
| WBC |
x103 cells/μL |
0.050–2.000 >2.000–100.000 >100.000–400.000 |
| RBC |
x106 cells/μL |
0.005–8.500 |
| HGB |
g/dL |
0.10–25.50 |
| MCV |
fL |
50.00–150.00 |
| RDW |
% |
10.00–40.00 |
| RDW-SD |
fL |
15.00–150.00 |
| PLT |
x103 cells/μL |
3.0–3000.0 |
| MPV |
fL |
5.00–25.00 |
|
|
|
- 50,000 patient results with graphics for standalone
- 100,000 patient results with graphics for power computer
|
|
|
QC with Levy-Jennings graph; XB/XM for moving averages, daily check, intelligent quality monitoring, MRV, IRF, customizable calibration and QC reminders and alerts, auto export of QC |
|
|
- Capacity: 20 five-tube cassettes
- Bar codes: digital bar code with 2D bar code capability
- Sample ID: up to 22 characters
|
|
|
PROService remote management system: maximum uptime with walk-away troubleshooting and preemptive alerts of instrument issues |
|
|
- Width: 75.57 cm
- Height: 174 cm, cover closed; 194 cm, cover lifted
- Depth: 82.80 cm, including removable back panel; 79.25 cm, excluding removable back panel
- Weight: 254 kg (560 lbs)
|
|
|
3.8 cm behind instrument for sufficient cooling; 15.2 cm per side of instrument for troubleshooting |